Description
Technical Features
Optics
Halogen quartz lamp 6V-10W
Photometric Range: -0.5 – 3.5 Abs
9 Interference filters positions: 340, 405, 450, 505, 550, 620 and
750 nm
2 free positions available.
Detection System
Photodiode 320-1000 nm
Aspiration System
Peristaltic pump driven by stepper motor.
Programmable volumen up to 500 ul.
Carry over <2% for volume > 400ul.
Automatic cleaning cycle.
Double Reading System
18 ul-quartz flow cell, termostatized by Peltier.
Standard holder for macro or semi-micro cuvettes 10 x 10 x 45 mm.
Temperature Control
Adjustable to 25º, 30º and 37ºC. Stability +/- 0.1ºC
Measurement Mode
End Point with sample or reagent blank.
Kinetics.
Two Points Kinetics.
Mono- and bichromatic measurements.
Reaction graphics.
Linearity checks.
Calibration Mode
Standard or Factor.
Factor storage in all methods.
Nonlinear curve fitting. Up to 10 standard points.
Automatic adjustment.
QC
10 Controls/test.
L. Jennings Plots.
High/low flags.
Operator Interface
25 functions on Display.
Thermal printer, 24 characters / line graphics.
20 x 20 characters graphic LCD with backlight.
Languages
English
Spanish
Portuguese
Installation Requirements
• Electrical Features
100-240 VAC 50/60 Hz
150 W
• Dimensions
32 x 17 x 38 cm
Weight: 6 kg
Certifications
CE/ ANMAT
ISO 9001